- All the pumps are fitted with air/water separator, except for item 021360-021350
- Control panels designed for two or more pumps are fitted with an electronic device to start/stop one
or more pumps automatically according to aspiration demand. - The connecting device for the electronic circuit of all electronic control panels (i.e. all control
panels marked by a final E, like C75/1E, C75/E etc.) must be completed with a peripheral P.B. unit
to be placed next to the tip support. - Tip supports and electropneumatic valves, which are necessary in all wet plants, are listed in the
dry plants section at pages 36 and 40. - Each suction machine can service a variable number of dental units depending on different
parameters: diameter and length of the piping, number of bends, brand of the dental unit
(requirements concerning aspiration can vary) therefore the number of dental units we suggest
should not be considered as binding. - The water feeding to the pumps PAL 28 and PAL 50 is made according to UNI EN 1717, November
2002, cat.5, protection unit A and B regulations (air break to drain). Water feedings to the pumps
PAL 14 and PAL 38 has to be made according to UNI EN 1717, November 2002, cat. 5 , protection
unit A and B regulations.
We suggest to read paragraph “The details that matter” before purchasing any Cattani equipment.
- Sound pressure level tested according to ISO regulation 3746 -1979 (E).
Parameters: r =1.5 – background noise < 34 dB (A) – instrument: Brüel & Kjær type 2232.
Output power single-phase 0.8 kW – 5
A, maximum flow 500 l/min, maximum
operating head for continuous service
2700 mm H2O. Suggested for 2 surgeries
working at the same time with surgical
tips or 1 surgery with tip no.17 and the
other with a surgical tip. Average water
consumption: 2 l/min with 2 bar piping
pressure. Sound pressure level with
canalized air …… dB (A).
W= 530 mm D= 340 mm H= 365 mm
Net Weight= 25 kg Gross Weight= 30 kg
The spittoon draining unit must be fitted
to the dental units in case the amalgam
separator is installed in a separate plant
room.
Components: spittoon filter, electrovalve
with magnet, branching tube for spittoon
draining unit, aspiration pipe and printed
board.
(Pneumatic tip support and
electropneumatic valve are not included)
Output power single-phase 0.8 kW – 5
A, maximum flow 500 l/min, maximum
operating head for continuous service
2700 mm H2O. Suggested for 2 surgeries
working at the same time with surgical
tips or 1 surgery with tip no.17 and the
other with a surgical tip. Average water
consumption: 2 l/min with 2 bar piping
pressure.
Sound pressure level with canalized air
56 dB (A).
W= 510 mm D= 360 mm H= 445 mm
Net Weight= 30 kg Gross Weight= 35 kg
The end-of-line valve sets into operation
to boost the air inlet and to empty
any siphon hindering the regular and
constant suction.
Components: centrifuge, 1 Hydrocyclone,
pressure relief valves, control panel.
Output power 0.25 kW – 2.8 A. Tested by
TÜV, according to ISO 11143 standard.
W= 470 mm D= 400 mm H= 590 mm
Output power three-phase 1.5 kW – 6.4/3.7 A,
maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O. Suggested for 4 surgeries
working at the same time with surgical
tips or two surgeries with surgical
tips and the other two with tips no.17.
Average water consumption 6 l/min with
4 bar piping pressure.
Sound pressure level with canalized air
68 dB (A).
W= 360 mm D= 310 mm H= 415 mm
Net Weight= 27 kg Gross Weight= 33 kg
Output power single-phase 1.3 kW –
9 A, maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O. Suggested for 4 surgeries
working at the same time with surgical
tips or two surgeries with surgical
tips and the other two with tips no.17.
Average water consumption 6 l/min with
4 bar piping pressure.
Sound pressure level with canalized air
68 dB (A).
W= 360 mm D= 310 mm H= 415 mm
Net Weight= 27 kg Gross Weight= 33 kg
W= 315 mm D= 175 mm H= 400 mm.
- Item 021391 is necessary to complete item 021360-021350.
Output power single-phase 1.3 kW –
9 A, maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O. Suggested for 4 surgeries
working at the same time with surgical
tips or two surgeries with surgical
tips and the other two with tips no.17.
Average water consumption 6 l/min with
4 bar piping pressure.
Sound pressure level with canalized air
68 dB (A).
W= 673 mm D= 549 mm H= 550 mm.
Net Weight= 42 kg Gross Weight= 53 kg
Output power three-phase 1.5 kW – 6.4/3.7 A,
maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O. Suggested for 4 surgeries
working at the same time with surgical
tips or two surgeries with surgical
tips and the other two with tips no.17.
Average water consumption 6 l/min with
4 bar piping pressure.
Sound pressure level with canalized air
68 dB (A).
W= 673 mm D= 549 mm H= 550 mm.
Net Weight= 42 kg Gross Weight= 53 kg
Output power three-phase 1.5 kW – 6.4/3.7 A,
maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O. Suggested for 4 surgeries
working at the same time with surgical
tips or two surgeries with surgical
tips and the other two with tips no.17.
Average water consumption 6 l/min with
4 bar piping pressure.
Sound pressure level with canalized air
64 dB (A).
W= 673 mm D= 435 mm H= 550 mm.
Net Weight= 45 kg Gross Weight= 56 kg
Output power single-phase 1.3 kW –
9 A, maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O. Suggested for 4 surgeries
working at the same time with surgical
tips or two surgeries with surgical
tips and the other two with tips no.17.
Average water consumption 6 l/min with
4 bar piping pressure.
Sound pressure level with canalized air
64 dB (A).
W= 673 mm D= 435 mm H= 550 mm.
Net Weight= 45 kg Gross Weight= 56 kg
Output power three-phase 1.5 kW – 6.4/3.7 A,
maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O. Suggested for 4 surgeries
working at the same time with surgical
tips or two surgeries with surgical
tips and the other two with tips no.17.
Average water consumption 6 l/min with
4 bar piping pressure.
Sound pressure level with canalized air
64 dB (A).
W= 673 mm D= 549 mm H= 550 mm.
Net Weight= 55 kg Gross Weight= 67 kg
Output power single-phase 1.3 kW –
9 A, maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O. Suggested for 4 surgeries
working at the same time with surgical
tips or two surgeries with surgical
tips and the other two with tips no.17.
Average water consumption 6 l/min with
4 bar piping pressure.
Sound pressure level with canalized air
64 dB (A).
W= 673 mm D= 549 mm H= 550 mm.
Net Weight= 55 kg Gross Weight= 67 kg
For two PAL 38 single-phase
W= 400 mm D= 220 mm H= 580 mm.
For two PAL 38 three-phase
W= 400 mm D= 220 mm H= 580 mm.
Output power single-phase 0.8 kW – 5
A, maximum flow 500 l/min, maximum
operating head for continuous service
2700 mm H2O.Suggested for 2 surgeries
working at the same time with surgical
tips or for 1 surgery with one tip no.17
and one surgical tip. Average water
consumption: 1.5 l/min with 2 bar piping
pressure. Sound pressure level with
canalized air 72 dB (A).
W= 530 mm D= 435 mm H= 515 mm
Net Weight= 35 kg Gross Weight= 40 kg
Output power single-phase 0.8 kW – 5
A, maximum flow 500 l/min, maximum
operating head for continuous service
2700 mm H2O.Suggested for 2 surgeries
working at the same time with surgical
tips or for 1 surgery with one tip no.17
and one surgical tip. Average water
consumption: 1.5 l/min with 2 bar piping
pressure.
Sound pressure level with canalized air 57 dB
(A). W= 530 mm D= 435 mm H= 560 mm.
Net Weight= 40 kg Gross Weight= 45 kg
For all pumps PAL 22 and 50 V.E.
The peripheral P.B. unit is required only
with electronic control panels (those
marked by a final E, like C75/1E, C75/E).
Each tip support has to be fitted with
one peripheral P.B. unit which must be
placed next to it.
W= 70 mm P= 70 mm H= 22 mm.
Pumps PAL 38 and PAL 50 V.E. cannot be
installed in parallel without the one-way
valve
For cold climates and outdoor
installation, on request the machine
can be fitted with an antifreeze device.
Complete with heating fan and
thermostat.
It comes into operation at 15 °C.
Output power, three-phase 1.5 kW – 6.4/3.7 A,
maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O.
Suggested for 4 surgeries working at
the same time with surgical tips or 2
surgeries with tips no 17 and the other
two with surgical tips.
Average water consumption 1.1 l/min
with 4 bar piping pressure.
Sound pressure level with canalized air
69 dB (A).
W= 673 mm D= 435 mm H= 550 mm
Net Weight= 48 kg Gross Weight= 60 kg
Output power single-phase 1.3 kW –
9 A, maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O.
Suggested for 4 surgeries working at
the same time with surgical tips or 2
surgeries with tips no 17 and the other
two with surgical tips.
Average water consumption 1.1 l/min
with 4 bar piping pressure.
Sound pressure level with canalized air
69 dB (A).
W= 673 mm D= 435 mm H= 550 mm
Net Weight= 48 kg Gross Weight= 60 kg
Output power three-phase 1.5 kW – 6.4/3.7 A,
maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O.
Suggested for 4 surgeries working at
the same time with surgical tips or 2
surgeries with tips no 17 and the other
two with surgical tips.
Average water consumption 1.1 l/min
with 4 bar piping pressure.
Sound pressure level with canalized air
69 dB (A).
W= 673 mm D= 549 mm H= 550 mm.
Net Weight= 52 kg Gross Weight= 65 kg
Output power single-phase 1.3 kW –
9 A, maximum flow 833 l/min, maximum
operating head for continuous service
3300 mm H2O.
Suggested for 4 surgeries working at
the same time with surgical tips or 2
surgeries with tips no 17 and the other
two with surgical tips.
Average water consumption 1.1 l/min
with 4 bar piping pressure.
Sound pressure level with canalized air
69 dB (A).
W= 673 mm D= 549 mm H= 550 mm.
Net Weight= 52 kg Gross Weight= 65 kg
For three PAL 50 V.E three-phase
W= 550 mm D= 250 mm H= 700 mm.
For two PAL 50 V.E single-phase
W= 400 mm D= 220 mm H= 580 mm.
For two PAL 50 V.E three-phase
W= 400 mm D= 220 mm H= 580 mm.
For three PAL 50 V.E single-phase
W= 550 mm D= 250 mm H= 700 mm.